Your browser doesn't support javascript.
loading
Part II: Effect of different evaluation methods to the application of a computer-aided prostate MRI detection/diagnosis (CADe/CADx) device on reader performance.
Maki, Jeffrey H; Patel, Nayana U; Ulrich, Ethan J; Dhaouadi, Jasser; Jones, Randall W.
  • Maki JH; Department of Radiology, University of Colorado Anschutz Medical Center, 12401 E 17th Ave (MS L954), Aurora, CO 80045, USA. Electronic address: jeffrey.maki@cuanschutz.edu.
  • Patel NU; University of New Mexico Department of Radiology, Albuquerque, NM, USA.
  • Ulrich EJ; BOT IMAGE, Inc., Omaha, NE, USA.
  • Dhaouadi J; BOT IMAGE, Inc., Omaha, NE, USA.
  • Jones RW; BOT IMAGE, Inc., Omaha, NE, USA.
Article en En | MEDLINE | ID: mdl-38702282
ABSTRACT

INTRODUCTION:

The construction and results of a multiple-reader multiple-case prostate MRI study are described and reported to illustrate recommendations for how to standardize artificial intelligence (AI) prostate studies per the review constituting Part I1.

METHODS:

Our previously reported approach was applied to review and report an IRB approved, HIPAA compliant multiple-reader multiple-case clinical study of 150 bi-parametric prostate MRI studies across 9 readers, measuring physician performance both with and without the use of the recently FDA cleared CADe/CADx software ProstatID.

RESULTS:

Unassisted reader AUC values ranged from 0.418 - 0.759, with AI assisted AUC values ranging from 0.507 - 0.787. This represented a statistically significant AUC improvement of 0.045 (α = 0.05). A free-response ROC (FROC) analysis similarly demonstrated a statistically significant increase in θ from 0.405 to 0.453 (α = 0.05). The standalone performance of ProstatID performed across all prostate tissues demonstrated an AUC of 0.929, while the standalone lesion level performance of ProstatID at all biopsied locations achieved an AUC of 0.710.

CONCLUSION:

This study applies and illustrates suggested reporting and standardization methods for prostate AI studies that will make it easier to understand, evaluate and compare between AI studies. Providing radiologists with the ProstatID CADe/CADx software significantly increased diagnostic performance as assessed by both ROC and free-response ROC metrics. Such algorithms have the potential to improve radiologist performance in the detection and localization of clinically significant prostate cancer.
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Año: 2024 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Año: 2024 Tipo del documento: Article