Development and Validation of a Simple High-Pressure Liquid Chromatography-Ultraviolet Detection Method for Simultaneous Quantitation of First-Line Anti-Tuberculosis Drugs in Formulations of Fixed-Dose Combination.
Vilvamani, Sudha; Mahalingam, Santhanamahalingam; Nhavilthodi, Sruthi; Murugesan, Dharman; Jeyakumar, Shanmugam Murugaiha.
  • Vilvamani S; Department of Clinical Pharmacology, ICMR-National Institute for Research in Tuberculosis, Chetpet, Chennai, Tamil Nadu 600031, India.
  • Mahalingam S; Department of Clinical Pharmacology, ICMR-National Institute for Research in Tuberculosis, Chetpet, Chennai, Tamil Nadu 600031, India.
  • Nhavilthodi S; Department of Clinical Pharmacology, ICMR-National Institute for Research in Tuberculosis, Chetpet, Chennai, Tamil Nadu 600031, India.
  • Murugesan D; Department of Clinical Pharmacology, ICMR-National Institute for Research in Tuberculosis, Chetpet, Chennai, Tamil Nadu 600031, India.
  • Jeyakumar SM; Department of Clinical Pharmacology, ICMR-National Institute for Research in Tuberculosis, Chetpet, Chennai, Tamil Nadu 600031, India.
J Chromatogr Sci ; 2024 May 08.
Article en En | MEDLINE | ID: mdl-38717231
ABSTRACT
The current treatment protocol for drug-sensitive tuberculosis involves all four first-line anti-tuberculosis drugs rifampicin, isoniazid, pyrazinamide and ethambutol hydrochloride in a single tablet, known as fixed-dose combination tablets. However, the analytical methods are scanty to test all these drugs simultaneously in a single run without any pre-sample process or using a simple method suitable for resource-limited settings. In this method, 50 mM potassium phosphate buffer containing 0.2% triethylamine (without pH adjustment) added with acetonitrile (982, v/v) was served as mobile phase A, while mobile phase B was 100% acetonitrile. All four drugs were separated within 10.3 min using a gradient mobile phase program in a C18 column (150 mm × 4.6 mm; 5 µm) and detected at two ultraviolet wavelengths (238 nm for rifampicin, isoniazid and pyrazinamide, and 210 nm for ethambutol hydrochloride). The method was selective, sensitive and linear with a correlation coefficient >0.999 with the acceptable precision and accuracy (<2% relative standard deviation) for all four drugs. In conclusion, the method is simple and it does not require any pH adjustment of the buffer/mobile phase, and within 11 min, the separation of all four drugs can be achieved. Overall, the method is suitable for quality testing of fixed-dose combination tablets in limited-resource settings.