Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation.
Watson, Jack Mitchell; Einsiedel, Paul F; Antippa, Phillip; Rangamuwa, Kanishka; Irving, Louis; Steinfort, Daniel P.
  • Watson JM; Department of Respiratory and Sleep Medicine, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Einsiedel PF; Department of Medical Imaging, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Antippa P; Cardiothoracic Surgery Unit, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Rangamuwa K; Department of Surgery, University of Melbourne, Melbourne, Victoria, Australia.
  • Irving L; Department of Respiratory and Sleep Medicine, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Steinfort DP; Department of Respiratory and Sleep Medicine, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
Respiration ; 103(8): 498-502, 2024.
Article en En | MEDLINE | ID: mdl-38885628
ABSTRACT

INTRODUCTION:

Endobronchial radiofrequency ablation (RFA) is a novel minimally invasive approach to management of peripheral non-small-cell lung cancer (NSCLC) in medically inoperable patients. Minimally invasive ablative techniques are generally delivered with cone-beam computed tomography (CBCT) guidance. CBCT requires a significant number of two dimensional imaging projections to be acquired which is then reconstructed as a three-dimensional cone-beam image. The objective of this study was to determine the radiation dosimetry consequent to use of CBCT guidance for bronchoscopic RFA.

METHODS:

Post hoc analysis of data following bronchoscopic RFA of stage I biopsy-confirmed NSCLC performed with CBCT. Effective dose estimates for these patients were calculated using PCXMC2.0 software.

RESULTS:

Ten patients underwent bronchoscopic RFA, with a median 3 (range 2-4) CBCT spins per procedure. Mean dose area product (DAP) per procedure was 7,778 µGy.m2 (±4,743) with an effective dose of 11.6 mSv (±7.4). The DAP per spin for these 10 patients varied from 83.8 to 8,625.6 µGy.m2 (effective dose range 0.15-13.81 mSv).

CONCLUSION:

This is the first study to report radiation dosimetry consequent to CT guidance for bronchoscopic RFA procedures. Effective doses appear comparable to other CT fluoroscopic procedures.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Broncoscopía / Carcinoma de Pulmón de Células no Pequeñas / Tomografía Computarizada de Haz Cónico / Neoplasias Pulmonares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Año: 2024 Tipo del documento: Article

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Broncoscopía / Carcinoma de Pulmón de Células no Pequeñas / Tomografía Computarizada de Haz Cónico / Neoplasias Pulmonares Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Año: 2024 Tipo del documento: Article