Your browser doesn't support javascript.
loading
Mass spectrometry coupling of chip-based supercritical fluid chromatography enabled by make-up flow-assisted backpressure regulation.
Weise, Chris; Fischer, Johannes; Belder, Detlev.
  • Weise C; Institute of Analytical Chemistry, University of Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany.
  • Fischer J; Institute of Analytical Chemistry, University of Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany.
  • Belder D; Institute of Analytical Chemistry, University of Leipzig, Linnéstrasse 3, 04103, Leipzig, Germany. belder@uni-leipzig.de.
Anal Bioanal Chem ; 416(20): 4447-4456, 2024 Aug.
Article en En | MEDLINE | ID: mdl-38907770
ABSTRACT
This work introduces a novel microfluidic backpressure pressure control developed for chip-based supercritical fluid chromatography (chipSFC). The presented on-chip pressure control mechanism involves the post-column addition of a viscous make-up stream, which enables pressure regulation within the range of 73 to 130 bar range. In contrast to approaches using mechanical backpressure regulators, this chip-based make-up-assisted pressure regulation offers a wear-free alternative that functions entirely through fluidic means and contributes minimally to extra column volume. It prevents phase separation of the supercritical mobile phase and, therefore, expands the analytical scope of chipSFC to detection systems with an ambient pressure interface. This was demonstrated by a proof-of-principle experiment, where a model mixture was separated within 30 s and detected using atmospheric pressure ionisation mass spectrometry.
Palabras clave