Your browser doesn't support javascript.
loading
Analysis of intra-fractional positioning correction performed by cone beam computed tomography in SBRT treatments.
García-Acilu, P; García Ruiz-Zorrilla, J; Hernando, O; Prado, A; Chen-Zhao, X; Montero, A; Martí, J; Zucca, D; Alonso, L; De la Casa, M A; Sánchez, E; Alonso, R; Rubio, C; Fernández-Letón, P.
  • García-Acilu P; Department of Medical Physics, University Hospital Madrid Puerta del Sur., Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: mgacilu@sescam.jccm.es.
  • García Ruiz-Zorrilla J; Department of Medical Physics, University Hospital Madrid Puerta del Sur., Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: jgruizzorrilla@hmhospitales.com.
  • Hernando O; Department of Radiation Oncology, University Hospital Madrid Puerta del Sur, Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: ohernando@hmhospitales.com.
  • Prado A; Department of Medical Physics, University Hospital Madrid Sanchinarro. C/de Oña, 10. 28050, Madrid, Spain. Electronic address: aprado@hmhospitales.com.
  • Chen-Zhao X; Department of Radiation Oncology, University Hospital Madrid Puerta del Sur, Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: xchen@hmhospitales.com.
  • Montero A; Department of Radiation Oncology, University Hospital Madrid Sanchinarro, C/de Oña, 10. 28050, Madrid, Spain. Electronic address: amontero@hmhospitales.com.
  • Martí J; Department of Medical Physics, University Hospital Madrid Puerta del Sur., Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: jmartiasenjo@hmhospitales.com.
  • Zucca D; Department of Medical Physics, University Hospital Madrid Sanchinarro. C/de Oña, 10. 28050, Madrid, Spain. Electronic address: dzucca@hmhospitales.com.
  • Alonso L; Department of Medical Physics, University Hospital Madrid Sanchinarro. C/de Oña, 10. 28050, Madrid, Spain. Electronic address: lalonso@hmhospitales.com.
  • De la Casa MA; Department of Medical Physics, University Hospital Madrid Sanchinarro. C/de Oña, 10. 28050, Madrid, Spain. Electronic address: mdelacasa@hmhospitales.com.
  • Sánchez E; Department of Radiation Oncology, University Hospital Madrid Puerta del Sur, Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: esanchezsaugar@hmhospitales.com.
  • Alonso R; Department of Radiation Oncology, University Hospital Madrid Puerta del Sur, Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain. Electronic address: rmalonso@hmhospitales.com.
  • Rubio C; Department of Radiation Oncology, University Hospital Madrid Puerta del Sur, Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain; Department of Radiation Oncology, University Hospital Madrid Sanchinarro, C/de Oña, 10. 28050, Madrid, Spain. Electronic address: crubio@hmhospitales.com.
  • Fernández-Letón P; Department of Medical Physics, University Hospital Madrid Puerta del Sur., Av. Carlos V, 70. 28938, Móstoles, Madrid, Spain; Department of Medical Physics, University Hospital Madrid Sanchinarro. C/de Oña, 10. 28050, Madrid, Spain. Electronic address: pfernandezleton@hmhospitales.com.
Phys Med ; 125: 104502, 2024 Sep.
Article en En | MEDLINE | ID: mdl-39216313
ABSTRACT

PURPOSE:

This study aims to evaluate the positioning correction extracted from Intra-fraction Cone Beam (IF-CBCT) images obtained during Stereotactic Body Radiotherapy (SBRT) treatments, and to assess whether its magnitude justifies its acquisition. In addition, the results obtained in lung, liver, and pancreas SBRTs with two deep inspiration breath-hold systems (DIBH), and for prostate with/without ultrasound (US) monitoring were compared.

METHODS:

1449 treatments, performed with two linear accelerators (LINACs) were retrospectively analyzed. DIBH were performed either with a spirometry-based device or a surface-guidance system and one LINAC was equipped with US monitoring system for prostate. Significance tests were used to account for differences between units.

RESULTS:

Group systematic error (M) was approximately -0.7 mm for DIBH treatments in superior-inferior (SI) direction with no difference (p > 0.7) between LINACs. Moreover, there was a SI difference of 0.5 mm for prostate treatments (p = 0.008), in favor of the US monitored one. In anterior-posterior (AP) direction, only liver treatments exhibited differences between LINACs, with the spirometer-based system being 0.8 mm inferior (p = 0.003). M<0.4 mm in left-right (LR) direction was found for all locations and LINACs. The spirometer-based system resulted in lower standard deviation of systematic and random errors in most components and locations, with a greater effect observed in liver SBRTs.

CONCLUSIONS:

The corrections made with IF-CBCT during SBRT treatments were not negligible. Both DIBH systems were effective in managing respiratory movements. However, the spirometry-based system was slightly more accurate. In addition, US monitoring of the prostate appeared to be useful in reducing target shift.