Chromatographic determination of N-acetyl-DL-tryptophan and octanoic acid in human albumin solutions.
J Chromatogr
; 333(2): 381-7, 1985 Oct 04.
Article
en En
| MEDLINE
| ID: mdl-4055947
Chromatographic procedures have been developed for determination of the stabilizers N-acetyl-DL-tryptophan and octanoic acid in human albumin solutions. N-Acetyl-DL-tryptophan and the internal standard, N-formyl-DL-tryptophan, were separated by liquid chromatography on a reversed-phase column with UV detection at 280 nm. Deproteinization and extraction were carried out with methanol. The extraction recovery at the level of 4.9 mM was 92.5 +/- 2.5% (S.D.) (n = 10), and the average coefficient of variation (C.V.) for replicate analyses of albumin solutions (mean = 2.57, 10.44 and 17.10 mM) was 1.10% (n = 27). Octanoic acid was determined gas chromatographically as its methyl ester, with nonanoic acid as the internal standard. The sample pretreatment included acidification, extraction with hexane and derivatization with methanol-sulphuric acid. The relative recovery from albumin solutions was 89.7 +/- 5.8% (S.D.) (n = 6), and replicate determinations of the compound yielded a C.V. of 5.5% (mean = 14.82 mM, n = 9).
Search on Google
Banco de datos:
MEDLINE
Asunto principal:
Triptófano
/
Caprilatos
/
Albúminas
Límite:
Humans
Idioma:
En
Año:
1985
Tipo del documento:
Article