Three-dimensional measurement of the mitral annulus by multiplane transesophageal echocardiography: in vitro validation and in vivo demonstration.
Legget, M E; Bashein, G; McDonald, J A; Munt, B I; Martin, R W; Otto, C M; Sheehan, F H.
  • Legget ME; Department of Medicine, Center of Bioengineering, University of Washington, Seattle 98195-6540, USA.
J Am Soc Echocardiogr ; 11(2): 188-200, 1998 Feb.
Article en En | MEDLINE | ID: mdl-9517558
ABSTRACT
Ten phantoms were scanned with a multiplane transesophageal echocardiographic probe in a water bath to assess a new method for three-dimensional modeling of the mitral annulus. The annulus was reconstructed from manually outlined borders with Fourier series in each of the three spatial coordinates. Comparisons with direct measurements by least-squares linear regression gave coefficients of determination of 0.99 for annular height, area, and circumference. Expressed as a percentage of their true values, the mean +/- SD of the errors were -0.1% +/- 3.0% for annular height, -2.8% +/- 3.1% for area, and -0.2% +/- 1.7% for circumference. The mean residual error length for phantoms was 0.64 mm compared with 1.21 mm in nine patients studied during general anesthesia. This method gives accurate and precise measurements of the mitral annulus in vitro and should be valuable for studying its morphology and dynamics in vivo.
Asunto(s)
Search on Google
Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Ecocardiografía Transesofágica / Válvula Mitral Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Año: 1998 Tipo del documento: Article
Search on Google
Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Ecocardiografía Transesofágica / Válvula Mitral Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Año: 1998 Tipo del documento: Article