A study of an independent dose verification software for brachytherapy / 中华放射肿瘤学杂志
Article en Zh | WPRIM | ID: wpr-799445
Biblioteca responsable: WPRO
ABSTRACT
Objective@#To report an implementation method and results of an independent brachytherapy dose verification software (DVS).@*Methods@#The DVS was developed based on Visual C+ + and the modular structure design was adopted. The DICOM RT files exported from the treatment planning system (TPS) were automatically loaded into the DVS. The TG-43 formalism was employed for dose calculation. Six cervical cancer patients who underwent brachytherapy were retrospectively selected to test the DVS. Different applicators were utilized for each patient. Dosimetric parameters and γ analysis (0.1cm, 5%) were used to evaluate the dose difference between the DVS and the TPS.@*Results@#Compared with the TPS dose, the γ pass rates of the doses calculated by the DVS were higher than 98%. For CTV, the dosimetric differences were less than 0.29% and 0.53% for D100% and D90%. For bladder, rectum and sigmoid, the agreement of D0.1cm3, D1cm3 and D2cm3 within a 0.5% level.@*Conclusion@#With minimal human-computer interactions, the DVS can verify the accuracy of dose calculated by TPS for brachytherapy.