Comparación de la estimación de gasto cardiaco por ecocardiografía versus biorreactancia y monitor ultrasónico de gasto cardiaco / Comparison of cardiac output estimation by echocardiography, bioreactance and ultrasonic cardiac output monitor / Comparação da estimativa do debito cardíaco pela ecocardiografia versus biorreatância e monitor ultrassônico
Cuesta Torres, Julia; Monares Zepeda, Enrique; Cruz López, Jorge; Cabrera Miranda, Luis; Barrón, Clemente; Galindo Martín, Carlos.
Afiliação
  • Cuesta Torres, Julia; Hospital San Ángel Inn Universidad. Ciudad de México. MX
  • Monares Zepeda, Enrique; Hospital San Ángel Inn Universidad. Ciudad de México. MX
  • Cruz López, Jorge; Hospital San Ángel Inn Universidad. Ciudad de México. MX
  • Cabrera Miranda, Luis; Hospital San Ángel Inn Universidad. Ciudad de México. MX
  • Barrón, Clemente; Hospital San Ángel Inn Universidad. Ciudad de México. MX
  • Galindo Martín, Carlos; Hospital San Ángel Inn Universidad. Ciudad de México. MX
Med. crít. (Col. Mex. Med. Crít.) ; 33(3): 116-120, may.-jun. 2019. graf
Article em Es | LILACS-Express | LILACS | ID: biblio-1154795
Biblioteca responsável: MX1.1
RESUMEN
Resumen En terapia intensiva uno de los monitoreos más frecuentes fundamentales para la toma de decisiones clínicas en la reanimación del choque séptico es la estimación del gasto cardiaco. La ecocardiografía se ha posicionado como el estándar de oro en monitoreo no invasivo del gasto cardiaco; sin embargo, este tipo de monitoreo es operador dependiente y debe ser realizado por un experto. Otras técnicas, como la biorreactancia y USCOM, son 100% no invasivas, y estas mediciones no dependen del operador ni del experto.

Objetivo:

Analizar la concordancia del gasto cardiaco por ecocardiografía versus biorreactancia y USCOM. Material y

métodos:

Estudio observacional prospectivo en el que se midió en 26 pacientes diagnosticados con shock séptico ingresados en la Unidad de Cuidados el gasto cardiaco por ecocardiografía, biorreactancia y sistema USCOM. El método de Bland Altman se realizó para analizar el acuerdo entre las diferentes técnicas. También se calculó el coeficiente de Linn en los mismos grupos de mediciones.

Resultados:

La biorreactancia frente a la ecocardiografía informó un sesgo (media de las diferencias) de -0.08 con una desviación estándar (precisión) de 0.85 y un intervalo de confianza de 95% de 1.59 a -1.75, finalmente un error de 24% y un coeficiente de Linn de 0.78. USCOM versus ecocardiografía informó un sesgo (media de las diferencias) de 1.11 con una desviación estándar (precisión) de 0.95 y un intervalo de confianza de 95% de 2.98 a -0.75, finalmente un error de 41% y un coeficiente de Linn de 0.6.

Conclusiones:

La biorreactancia es un prometedor monitoreo continuo al parecer comparable a la ecocardiografía, por otro lado el monitoreo USCOM no es equiparable a ecocardiografía o biorreactancia. Se necesitan más estudios para comprobar esta observación.
ABSTRACT
Abstract Cardiac output is the most frequent advanced monitoring in critical care units given its utility to guide resuscitation in septic shock. Echocardiography has been postulated has the gold standard for non-invasive cardiac output measurement, however this technique is operator-dependent and should be performed by an expert. Other techniques such as bioreactance and Doppler monitoring by semiautomatic USCOM system are 100% non-invasive techniques, also these measurements are not operator or expert dependent.

Objective:

To analyze the agreement of cardiac output by echocardiography vs Bioreactance and USCOM. Material and

methods:

Prospective observational study in which 26 patients diagnosed with septic shock admitted to the intensive care unit were cardiac output was measured by echocardiography, bioreactance and USCOM system. Bland Altman method was performed to analyze the agreement between the different techniques echocardiography vs USCOM and echocardiography vs bioreactance. Also Linn coefficient was calculated in the same groups of measurements.

Results:

Bioreactance vs echocardiography reported a bias (mean of the differences) of -0.08 with a standard deviation (precision) of 0.85 and a 95% confidence interval of 1.59 to -1.75, finally an error of 24% and Linn coefficient of 0.78. USCOM vs echocardiography reported a bias (mean of the differences) of 1.11 with a standard deviation (precision) of 0.95 and a 95% confidence interval of 2.98 to -0.75, finally an error of 41% and Linn coefficient of 0.6.

Conclusions:

Bioreactance is a promising technique for cardiac output measurement given its good agreement with echocardiography, in the other hand USCOM did not showed a satisfactory agreement with echocardiography, more studies are needed to verify the clinical usefulness of these different techniques.
RESUMO
Resumo Na unidade de terapia intensiva, um dos monitoramentos mais frequentes essenciais para a tomada de decisão clínica na ressuscitação do choque séptico é a estimativa do débito cardíaco. A ecocardiografia tem sido posicionada como o «padrão ouro¼ na monitorização não invasiva do débito cardíaco, no entanto esse tipo de monitoramento é dependente do operador e deve ser executado por um especialista. Outras técnicas como a biorreatância e o USCOM, são técnicas 100% não invasivas, e essas medidas não dependem do operador ou do especialista.

Objetivo:

Analisar a concordância do débito cardíaco pela ecocardiografia versus Biorreatância e USCOM. Material e

métodos:

Estudo observacional prospectivo em que 26 pacientes com diagnóstico de choque séptico admitidos na unidade de terapia intensiva, onde mediram-se o débito cardíaco por ecocardiografia, biorreatância e sistema USCOM. O método de Bland Altman foi utilizado para analisar a concordância entre as diferentes técnicas. O coeficiente de Linn também foi calculado nos mesmos grupos de medidas.

Resultados:

A biorreatância versus ecocardiografia relatou uma tendência (diferenças médias) de -0.08 com um desvio padrão (precisão) de 0.85 e um intervalo de confiança de 95% de 1.59 a -1.75, finalmente um erro de 24% e um coeficiente de Linn de 0.78. USCOM versus ecocardiografia relatou uma tendência (média de diferenças) de 1.11 com um desvio padrão (precisão) de 0.95 e um intervalo de confiança de 95% de 2.98 a -0.75, finalmente um erro de 41% e um coeficiente de Linn de 0.6.

Conclusões:

Biorreatância é um monitoramento contínuo promissor aparentemente comparável ao ecocardiograma; por outro lado, o monitoramento por USCOM não é comparável à ecocardiografia ou biorreactância. Mais estudos são necessários para verificar essa observação.