Reagent injection FIA system for lead determination by hydride generation - quartz-tube atomic absorption spectrometry.
Anal Bioanal Chem
; 374(1): 115-9, 2002 Sep.
Article
em En
| MEDLINE
| ID: mdl-12207251
A method is proposed for the determination of lead by generation of its hydride and detection by quartz-tube AAS using a reagent injection FIA system based on the injection of sodium tetrahydroborate. Lead hydride generation was carried out using a combination of 0.5 M nitric acid, 10% m/ v hydrogen peroxide and 10% m/ v sodium tetrahydroborate. The characteristic concentration obtained was 3.1 ng mL(-1) and the detection limit was 2.6 ng mL(-1) for an injected volume of 0.125 mL of tetrahydroborate.