Your browser doesn't support javascript.
loading
Evaluation of the dosimetric response for CaSO4:Dy at low temperature.
Cruz-Zaragoza, E; Ramos-Bernal, S; Negrón-Mendoza, A; Azorin, J.
Afiliação
  • Cruz-Zaragoza E; Instituto de Ciencias Nucleares, UNAM, México, DF México. ecruz@nuclecu.unam.mx
Radiat Prot Dosimetry ; 100(1-4): 447-9, 2002.
Article em En | MEDLINE | ID: mdl-12382918
ABSTRACT
Homemade solid state CaSO4Dy detectors were tested to evaluate their response to gamma radiation at liquid nitrogen temperature (77 K). The dosemeters were irradiated with doses between 12 and 1071 Gy. For this study these dosemeters were exposed to gamma rays with a dose rate of 1.19 Gy.min(-1). The analysis for these crystals was made by thermoluminiscence. The dose response at liquid nitrogen temperature was linear in the dose range studied and it is about 20% lower with respect to the response at room temperature. The response is reproducible with the same geometric set-up.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Dosimetria Termoluminescente / Sulfato de Cálcio Tipo de estudo: Evaluation_studies Idioma: En Ano de publicação: 2002 Tipo de documento: Article