Synchrotron soft X-ray and field-emission electron sources: a comparison.
Spence, J C H; Howells, M R.
Afiliação
  • Spence JC; Department of Physics and Astronomy, Arizona State University, Tempe, AZ 85287-1504, USA. spence@asu.edu
Ultramicroscopy ; 93(3-4): 213-22, 2002 Dec.
Article em En | MEDLINE | ID: mdl-12492232
The soft X-ray spectral region and the useful range of electron energy-loss spectroscopy are very similar, both including the energy range 100-1000 eV. Moreover, well-developed monochromators and parallel detection devices with comparable resolution exist for both. Despite the differing interactions of electrons and photons, many complementary experiments in imaging, spectroscopy and diffraction have been performed using both techniques. We therefore compare the brightness, degeneracy, monochromaticity, beam size, source size, spatial and temporal coherence of field-emission electron beams and soft X-ray synchrotron radiation from typical undulators. Recent brightness values for nanotip field emitters and undulators, both measured and calculated, are provided with examples from the Advanced Light Source synchrotron-radiation facility at Berkeley USA. The quantum mechanical upper limit on source brightness, as well as relationships among beam brightness, coherence parameters, and degeneracy, are discussed. Factors which limit these parameters and methods of measurement are reviewed, and the implications for diffraction, imaging and spectroscopic experiments as well as radiation damage are briefly commented on.
Buscar no Google
Base de dados: MEDLINE Idioma: En Ano de publicação: 2002 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Idioma: En Ano de publicação: 2002 Tipo de documento: Article