Your browser doesn't support javascript.
loading
Quartz-crystal microbalance immunosensor for Schistsoma-japonicum-infected rabbit serum.
Wu, Zhao-Yang; Shen, Guo-Li; Wang, Shi-Ping; Yu, Ru-Qin.
Afiliação
  • Wu ZY; State Key Laboratory of Chemo/Biosensing and Chemometrics, Institute of Chemistry and Chemical Engineering, Hunan University, Changsha 410082, China. zywu@hnu.net.cn
Anal Sci ; 19(3): 437-40, 2003 Mar.
Article em En | MEDLINE | ID: mdl-12675355
ABSTRACT
A quartz-crystal microbalance immunosensor (QCM) has been developed for the direct determination of Schistosoma-japonicum-infected rabbit serum. A self-assembled monolayer with carboxyl groups was first coated on a gold electrode of a quartz-crystal resonator by the spontaneous adsorption of 3-mercaptopropionic acid. Schistosoma-japonicum molecular antigen of 32 kD molecular weight was then covalently attached to the crystal surface. The QCM immunosensor was used to detect infected rabbit serum (IRS49-2000); a maximum titer of 1800 was achieved.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Quartzo / Schistosoma japonicum / Esquistossomose / Técnicas Biossensoriais Limite: Animals Idioma: En Ano de publicação: 2003 Tipo de documento: Article