Inductively coupled plasma mass spectrometry (ICP-MS), highly sensitive inorgnic analytic technique, fits to determine ultra-trace rare-earth elements in human plasma. Under the optimized conditions detection limits for 15 rare-earth elements are in the range of 0.7 (for Eu)-5.4 (for Gd) ng.L-1. Indium as an internal standard element is used to compensate for matrix suppression effect and sensitivity drift. Three kinds of preparation methods, diluted with 1% HNO3, digested with HNO3-H2O2 and with HNO3-HClO4, are checked and compared, and the former is the simplest way to be measured. The samples diluted with 1% HNO3, stored in 4 degrees C, are very steady for 16 days. With the method, 11 healthy plasma samples in Changchun area of China are analysed.