Correspondence of Tono-Pen intraocular pressure measurements performed at the central cornea and mid-peripheral cornea.
Sullivan-Mee, Michael; Pham, Fortsmith.
Afiliação
  • Sullivan-Mee M; Albuquerque VA Medical Center, Eye Clinic, Albuquerque, New Mexico 87108, USA. Michael.Sullivan-Mee@med.va.gov
Optometry ; 75(1): 26-32, 2004 Jan.
Article em En | MEDLINE | ID: mdl-14717278
ABSTRACT

BACKGROUND:

As the awareness of the influence of central corneal thickness (CCT) on Goldmann tonometry has increased, many publications have questioned the accuracy of Goldmann intraocular pressure (IOP) measurement. The Tono-Pen, because it indents a much smaller surface area when compared to a Goldmann probe, may be less affected by corneal thickness variations when compared with Goldmann tonometry.

METHODS:

Forty human subjects with no history of refractive surgery participated in this study. The IOP of the right eye of each subject was measured with the Goldmann tonometer, the Tono-Pen at the central cornea, and the Tono-Pen at the mid-peripheral cornea. An ultrasonic DGH Pachette pachymeter was used to measure the central and mid-peripheral corneal thickness at the location of the IOP readings.

RESULTS:

Tono-Pen measurements at the central and mid-peripheral cornea highly correlated (r = 0.933), and did not significantly differ (p = 0.646). The IOP readings with the Goldmann tonometer (r= 0.406), the Tono-Pen at the central cornea (r = 0.453), and the Tono-Pen at the mid-peripheral cornea (r = 0.321) showed a positive correlation to corneal thickness. The Goldmann and Tono-Pen tonometers differed significantly in the measurement of IOP at the central cornea (p = 0.007), but were positively correlated (r = 0.674).

CONCLUSIONS:

The Tono-Pen IOP measurement at the central cornea highly approximated Tono-Pen IOP measurement at the mid-peripheral cornea. Furthermore, although not highly correlated, both the Goldmann and Tono-Pen tonometers showed a significantly positive correlation between IOP and corneal thickness measurements.