Your browser doesn't support javascript.
loading
Standardization of 65Zn by 4piPC-gamma coincidence counting method with efficiency extrapolation.
Sahagia, Maria; Ivan, C; Grigorescu, E L; Capogni, M; De Felice, P; Fazio, A.
Afiliação
  • Sahagia M; National Institute of R&D for Physics and Nuclear Engineering Horia Hulubei, IFIN-HH, POB MG-6, Bucharest RO-76900, Romania. msahagia@ifin.nipne.ro
Appl Radiat Isot ; 60(2-4): 423-7, 2004.
Article em En | MEDLINE | ID: mdl-14987679
A 65Zn solution was standardized by the 4piPC-gamma efficiency-extrapolation coincidence counting method. Theoretical aspects of coincidence equations, efficiency equations and linearity conditions are reviewed. Experimental measurements were performed for two low level discrimination thresholds for the PC channel (counting K+L or K X-rays or Auger electrons from EC decay) and for three different settings of the gamma window. Requirements on gamma channel set-up for linear extrapolation were established by using a Pb absorber or by proper gamma window setting. The measured activity values were discussed and found in good agreement with those obtained with a calibrated ionization chamber.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Padrões de Referência / Espectrometria gama / Radioisótopos de Zinco / Algoritmos / Raios gama Tipo de estudo: Diagnostic_studies / Evaluation_studies Idioma: En Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google