Your browser doesn't support javascript.
loading
Pacing in high field cardiac magnetic resonance imaging:.
Nahrendorf, Matthias; Hiller, Karl-Heinz; Hu, Kai; Zeijlemaker, Volkert; Griswold, Mark; Bauer, Wolfgang R.
Afiliação
  • Nahrendorf M; Medizinische Universitätsklinik Würzburg, Wuerzburg, Germany. M.Nahrendorf@Medizin.uni-wuerzburg.de
Pacing Clin Electrophysiol ; 27(5): 671-4, 2004 May.
Article em En | MEDLINE | ID: mdl-15125728
ABSTRACT
Currently, cardiac MRI is contraindicated in patients with an implanted pacemaker or ICD due to safety hazards. However, MRI is promising to play a key role in cardiac diagnostics in near future. This study examined a rat with an implanted pacemaker pacing at a rate of 460/min with high field cardiac MRI. This study showed that pacing during cardiac imaging at 7 Tesla was possible. The pacemaker program was not disturbed by the high field or the strong gradients (maximum dB/dt 400 mT/s). The only noticeable effect on the MRI signal was a signal void of 2 cm around the device.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Imageamento por Ressonância Magnética Limite: Animals Idioma: En Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Imageamento por Ressonância Magnética Limite: Animals Idioma: En Ano de publicação: 2004 Tipo de documento: Article