Determination of ephedrine and related compounds in pharmaceutical preparations by ion chromatography with direct conductivity detection.
Biomed Chromatogr
; 19(4): 266-71, 2005 May.
Article
em En
| MEDLINE
| ID: mdl-15651083
An ion chromatographic method with conductivity detection for the simultaneous determination of ephedrine, pseudoephedrine and norephedrine was developed. A mixture of 2.0 mmol/L HNO3 and 2% (v/v) acetonitrile was used as eluent. The three ephedrine-like compounds were separated and determined within 20 min. The linear ranges were 0.08-50 microg/mL for ephedrine, 0.08-40 microg/mL for pseudoephedrine and 0.06-40 microg/mL for norephedrine. The detection limits were 0.03 microg/mL for ephedrine and pseudoephedrine, and 0.02 microg/mL for norephedrine. The method has been applied successfully to the determination of these sympathomimetics in pharmaceutical preparations and in Ephedra herbs.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Preparações Farmacêuticas
/
Cromatografia por Troca Iônica
/
Efedrina
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Ano de publicação:
2005
Tipo de documento:
Article