Your browser doesn't support javascript.
loading
[Valve replacement with the Omnicarbon valve prosthesis. A 10-year follow-up]. / Troca de valva com a prótese valvar modelo Omnicarbon. Seguimento clínico de 10 anos.
Florez, Santiago; Di Stefano, Salvatore; Carrascal, Yolanda; Bustamante, Juan; Fulquet, Enrique; Echevarria, José Ramón; Elena, Casquero; Gualis, Javier; Fiz, Luis.
Afiliação
  • Florez S; Instituto de Ciencias del Corazon (ICICOR), Servicio de Cirugía Cardíaca, Hospital Universitario de Valladolid, Spain.
Arq Bras Cardiol ; 84(5): 371-5, 2005 May.
Article em Pt | MEDLINE | ID: mdl-15917968
ABSTRACT

OBJECTIVE:

We retrospectively examined the outcomes of 264 patients who underwent consecutive Omnicarbon valve implantation surgery between April 1985 and May 1995.

METHODS:

At the time of surgery, patients who received this mechanical prosthesis averaged 57+/-11 years of age. Omnicarbon valves were placed in the aortic position in 36% of the cases, in the mitral position in 44%, and in both positions in 20%. Follow-up was carefully performed, with most patients undergoing physical examination at our clinic. While taking the case history, cardiac physicians specifically questioned the patient about valve-related complications.

RESULTS:

Accumulated total patient-years is 1291, with a mean follow-up time of 5.4 years. Survival at 10 years is 79.4+/-3.9%, including all causes of death and early mortality. Complications recorded during the 11-year study include thromboembolism (0.1%), hemorrhage (0.4%), endocarditis (0.2%), and nonstructural failure (1.2%). No hemolytic anemia, valve thrombosis, or structural failure was detected during this long-term experience. Functional capability of these patients was subjectively assessed by the NYHA classification system. With follow-up time averaging over 5 years, 97% of our Omnicarbon valve patients are in NYHA I or II.

CONCLUSION:

The Omnicarbon mechanical prosthesis provides a good clinical performance for up to 10 years in both the aortic and mitral positions. Results indicated a low incidence of thromboembolism and of hemorrhagic complications.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca / Doenças das Valvas Cardíacas / Valva Mitral Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: Pt Ano de publicação: 2005 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Valva Aórtica / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca / Doenças das Valvas Cardíacas / Valva Mitral Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: Pt Ano de publicação: 2005 Tipo de documento: Article