OSEM reconstruction, associated with temporal fourier and depth-dependant resolution recovery filtering, enhances results from sestamibi and 201Tl 16-interval gated SPECT.
Marie, Pierre-Yves; Djaballah, Wassila; Franken, Philippe R; Vanhove, Chris; Muller, Marc A; Boutley, Henri; Poussier, Sylvain; Olivier, Pierre; Karcher, Gilles; Bertrand, Alain.
Afiliação
  • Marie PY; Department of Nuclear Medicine, University Hospital, Nancy, France. py.marie@chu-nancy.fr
J Nucl Med ; 46(11): 1789-95, 2005 Nov.
Article em En | MEDLINE | ID: mdl-16269591
ABSTRACT
UNLABELLED Gated SPECT recorded with 16 intervals determines left ventricular (LV) ejection fraction more accurately than does gated SPECT recorded with 8 intervals but produces higher image noise. This study aimed to assess the results from sestamibi and (201)Tl 16-interval gated SPECT when both signal-to-noise ratio and spatial resolution were enhanced with an original method of reconstruction.

METHODS:

Forty patients with coronary artery disease underwent (201)Tl and sestamibi 16-interval gated SPECT and, to be used as a reference, cardiac MRI. Assessments of global and regional LV function provided by ordered-subsets expectation maximization (OSEM) with depth-dependant resolution recovery and temporal Fourier filtering were compared with those from conventional filtered backprojection (FBP) previously optimized by screening various filter frequencies and various temporal smoothing levels.

RESULTS:

For both tracers, LV ejection fraction was determined best when the association of OSEM with depth-dependant resolution recovery was used alone, with temporal Fourier filtering, or with a slight 2-frame temporal smoothing Mean absolute values of relative errors ranged from 3.2% to 3.6% (4.0%-7.9% for FBP), and coefficient correlation ranged from 0.91 to 0.93 (0.70-0.91 for FBP). Among these 3 reconstruction methods, the association of OSEM with depth-dependant resolution recovery with temporal Fourier filtering provided the highest signal-to-noise ratio, with mean increases of 54% for sestamibi and 80% for (201)Tl when compared with FBP, and the best analysis of segmental contractility, with exact agreement rates with MRI being 73% for (201)Tl and 79% for sestamibi.

CONCLUSION:

OSEM associated with temporal Fourier filtering and depth-dependant resolution recovery filtering enhances the LV function assessment provided by sestamibi and (201)Tl 16-interval gated SPECT and dramatically reduces image noise, a property that enhances and facilitates image interpretation.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Volume Sistólico / Tálio / Doença da Artéria Coronariana / Aumento da Imagem / Imagem do Acúmulo Cardíaco de Comporta / Tecnécio Tc 99m Sestamibi / Disfunção Ventricular Esquerda Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2005 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Volume Sistólico / Tálio / Doença da Artéria Coronariana / Aumento da Imagem / Imagem do Acúmulo Cardíaco de Comporta / Tecnécio Tc 99m Sestamibi / Disfunção Ventricular Esquerda Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2005 Tipo de documento: Article