A biomechanical study of the fatigue characteristics of thoracolumbar fixation implants in a calf spine model.
Wittenberg, R H; Shea, M; Edwards, W T; Swartz, D E; White, A A; Hayes, W C.
Afiliação
  • Wittenberg RH; Department of Orthopaedic Surgery, St. Josef Hospital, Ruhr University, Bochum, Germany.
Spine (Phila Pa 1976) ; 17(6 Suppl): S121-8, 1992 Jun.
Article em En | MEDLINE | ID: mdl-1631711
Clinical failures of internal fixation implants for the treatment of the thoracolumbar spine are generally attributed to fatigue. Few studies, however, have characterized changes in fixation rigidity with time or subjected spine-implant fixation constructs to fatigue loading until failure. Fatigue characteristics of five dorsally applied spinal fixation implants were determined using lumbosacral calf spines, with an L3 vertebrectomy, loaded cyclically in combined compression (maximum 605 N) and flexion (maximum 16 Nm) for up to 100,000 cycles. Displacement transducers monitored motion at the site of instability and at the segment above the implants. Flexibility and strain at these segments were then calculated. A one-way analysis of variance showed that there were no significant differences in flexibility of the five fixation constructs (P greater than .05). A multiple Bonferroni test revealed that the AO and Kluger fixateur interne and Steffee plates, with fixation at L2 and L4, allowed significantly more strain (P less than .01) across the site of instability than did Harrington rods and Luque plates with fixation at L1, L2, L4, and L5. There were no significant differences between fixation constructs in initial strain above the implants. After 10,000 cycles, however, there were significant increases in strain across the segment above the Luque and Harrington implants (P less than .05). Failure of the AO Schanz screw occurred in three of six constructs at a mean of 73,300 cycles. The Steffee screws failed in four of five constructs at a mean of 20,800 cycles. The rods of the Kluger fixateur interne broke in four of five constructs at a mean of 47,800 cycles, and one screw slipped at 11,000 cycles. There were no metal failures in the Harrington or Luque implants.(ABSTRACT TRUNCATED AT 250 WORDS)
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Vértebras Torácicas / Fixadores Internos / Vértebras Lombares Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 1992 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Vértebras Torácicas / Fixadores Internos / Vértebras Lombares Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 1992 Tipo de documento: Article