Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study.
Hatakeyama, Y; Kakeda, S; Ohnari, N; Moriya, J; Oda, N; Nishino, K; Miyamoto, W; Korogi, Y.
Afiliação
  • Hatakeyama Y; Department of Radiology, University of Occupational and Environmental Health School of Medicine, Kitakyushu, Japan.
AJNR Am J Neuroradiol ; 28(4): 645-50, 2007 Apr.
Article em En | MEDLINE | ID: mdl-17416814
ABSTRACT
BACKGROUND AND

PURPOSE:

Compared with image intensifier television (I.I.-TV) system, an angiography system using the flat panel detector (FPD) of direct conversion type has a high spatial resolution, which may improve image quality, reduce patient exposure, or both. Our purpose was to evaluate the detection of simulated aneurysmal blebs under dose reduction with the FPD system in comparison with the I.I.-TV system. MATERIALS AND

METHODS:

A vascular phantom was designed to simulate various intracranial aneurysms with and without blebs, and this phantom was filled with 3 different concentrations of contrast material (300, 150, and 100 mg I/mL). 2D digital subtraction angiography (DSA) at low-dose mode of FPD system was compared with 2D DSA at a standard-dose mode of FPD system and a conventional mode of I.I.-TV system. Data analysis was based on 171 observations (57 aneurysms [20 with bleb and 37 without bleb] x 3 contrast material concentrations) by each of 7 radiologists, and the detection performances of blebs were compared using a receiver operating characteristic (ROC) analysis.

RESULTS:

The mean dose measurements with a phantom during 2D DSA were 0.36 mGy/frame with low-dose mode of FPD system, 0.72 mGy/frame with standard-dose mode of FPD system and 0.76 mGy/frame with I.I.-TV system. The mean Az at 100 mg I/mL was significantly higher for low-dose mode of FPD than for conventional-dose mode of I.I.-TV mean Az, 0.85 versus 0.56; P < .01), though differences were not significant with 150 and 300 mg I/mL between both systems.

CONCLUSION:

The FPD system allows a considerable dose reduction during 2D DSA without loss of the image quality.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Angiografia Cerebral / Intensificação de Imagem Radiográfica / Angiografia Digital / Aneurisma Intracraniano Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2007 Tipo de documento: Article