Your browser doesn't support javascript.
loading
Optimization of the cutoff value for the Aspergillus double-sandwich enzyme immunoassay.
Maertens, Johan A; Klont, Rocus; Masson, Christine; Theunissen, Koen; Meersseman, Wouter; Lagrou, Katrien; Heinen, Christine; Crépin, Brigitte; Van Eldere, Johan; Tabouret, Marc; Donnelly, J Peter; Verweij, Paul E.
Afiliação
  • Maertens JA; Department of Hematology, Universitaire Ziekenhuizen, Catholic University Leuven, Leuven, Belgium. johan.maertens@uz.kuleuven.ac.be
Clin Infect Dis ; 44(10): 1329-36, 2007 May 15.
Article em En | MEDLINE | ID: mdl-17443470
ABSTRACT

BACKGROUND:

Many health care centers worldwide use the Platelia Aspergillus enzyme immunoassay (PA-EIA; Bio-Rad Laboratories) for diagnosis of invasive aspergillosis (IA). A cutoff optical density (OD) index of 1.5 was originally recommended by the manufacturer, but in practice, most institutions use lower cutoff values. Moreover, a cutoff OD index of 0.5 was recently approved in the United States. In the present study, we set out to optimize the cutoff level by performing a retrospective analysis of PA-EIA values for samples that had been obtained prospectively from adult patients at risk for IA at 2 European health care centers.

METHODS:

In total, 239 treatment episodes were included of which there were 19 episodes of proven IA and 19 episodes of probable IA. Per-episode and per-test analyses and receiver operating characteristic curves were used to determine the optimal cutoff value.

RESULTS:

In the per-episode analysis, lowering the cutoff OD index for positivity from 1.5 to 0.5 increased the overall sensitivity by 21% (from 76.3% to 97.4%) but decreased the overall specificity by 7% (from 97.5% to 90.5%). Requiring 2 consecutive samples with an OD index > or = 0.5 resulted in the highest test accuracy, with an improved positive predictive value. At a cutoff OD index of 0.5, the antigen test result was positive during the week before conventional diagnosis in 65% of cases and during the week of diagnosis in 79.5% of cases.

CONCLUSIONS:

A cutoff OD index of 0.5--identical to the approved cutoff in the United States--improves the overall performance of the PA-EIA for adult hematology patients.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Aspergilose / Aspergillus / Técnicas Imunoenzimáticas Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2007 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Aspergilose / Aspergillus / Técnicas Imunoenzimáticas Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2007 Tipo de documento: Article