Accuracy of the HoMedics BPA-300, a home blood pressure monitor using the auscultatory method.
Blood Press Monit
; 12(5): 339-43, 2007 Oct.
Article
em En
| MEDLINE
| ID: mdl-17890974
PURPOSE: To determine whether the HoMedics Model BPA-300 meets American Association for the Advancement of Medical Instrumentation (AAMI) accuracy standards. METHOD: Simultaneous auscultatory measurements by two observers using a mercury manometer were obtained in participants during measurement with the HoMedics Model BPA-300. Triple measurements were obtained after a minimum of 5 min of rest and 1 min between readings. RESULTS: In total, 132 participants had 396 nurse-measured blood pressures (BP) and 385 successful systolic and 383 diastolic monitor readings. The initial auscultation systolic BP was less than 100 mmHg in 17 participants (12.9%), between 100 and 160 mmHg in 108 participants (81.8%), and greater than 160 mmHg in seven (5.3%) of participants. The average difference between the observers and device was 0.6+/-4.3 mmHg for systolic and 0.5+/-4.3 mmHg for diastolic-measured BPs by AAMI method 1, which met this guideline. The average difference between observers and device by AAMI method 2 was 0.9+/-4.0 mmHg for systolic and 0.5+/-3.9 mmHg for diastolic-measured BP, which met this guideline. CONCLUSION: The HoMedics Model BPA-300 meets AAMI accuracy standards using methods 1 and 2.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Auscultação
/
Pressão Sanguínea
/
Monitores de Pressão Arterial
/
Monitorização Ambulatorial da Pressão Arterial
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
Limite:
Adult
/
Aged
/
Female
/
Humans
/
Male
/
Middle aged
País como assunto:
America do norte
Idioma:
En
Ano de publicação:
2007
Tipo de documento:
Article