An electrically-activated dynamic tissue-equivalent phantom for assessment of diffuse optical imaging systems.
Hebden, Jeremy C; Brunker, Joanna; Correia, Teresa; Price, Ben D; Gibson, Adam P; Everdell, N L.
Afiliação
  • Hebden JC; Department of Medical Physics & Bioengineering, University College London, Gower Street, London WC1E 6BT, UK.
Phys Med Biol ; 53(2): 329-37, 2008 Jan 21.
Article em En | MEDLINE | ID: mdl-18184989
ABSTRACT
A novel design of solid dynamic phantom with tissue-like optical properties is presented, which contains variable regions of contrast which are activated electrically. Reversible changes in absorption are produced by localized heating of targets impregnated with thermochromic pigment. A portable, battery-operated prototype has been constructed, and its optical and temporal characteristics have been investigated. The phantom has been developed as a means of assessing the performance of diffuse optical imaging systems, such as those used to monitor haemodynamic changes in the brain and other tissues. Images of the phantom have been reconstructed using data acquired with a continuous wave optical topography system.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imagens de Fantasmas / Tomografia Óptica / Imagem Corporal Total Idioma: En Ano de publicação: 2008 Tipo de documento: Article