Evaluation of a commercial radiochromatography module as an arterial blood activity monitor.
Phys Med Biol
; 53(2): 339-51, 2008 Jan 21.
Article
em En
| MEDLINE
| ID: mdl-18184990
Input functions required for positron emission tomography (PET) tracer kinetic modeling are often obtained from arterial blood. In some situations, using short-lived radiotracers, e.g. [(15)O]water, rapid sample handling is required. A method used at several facilities is to pump blood through a detector system at a constant rate. We investigate the suitability of a commercial radiochromatography module (IN/US Posi-RAM) for this new use. The Posi-RAM consists of two 2.5 cm (length) x 2.5 cm (diameter) cylindrical bismuth germanate (BGO) detectors that can operate in coincidence mode. Arterial blood is transported through the system via a length of tubing with flow rate controlled by a peristalsis pump. A custom-counting loop and support frame were designed for the Posi-RAM for PET studies. System sensitivity was determined to be 1.1 x 10(4) cps/(MBq ml(-1)). Dead time as a function of count-rate was found to be less than 1% for concentrations below 3.5 MBq ml(-1), a range encompassing all human-study values. In a human study, the performance of the device was found to be similar to that of the facility's current blood monitor (Siemens Fluid Monitor). We conclude that the Posi-RAM has the necessary sensitivity and count-rate capabilities to be used as a real-time blood activity monitor.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Artérias
/
Radiometria
/
Sangue
/
Análise Química do Sangue
/
Processamento de Sinais Assistido por Computador
/
Cromatografia
/
Compostos Radiofarmacêuticos
/
Tomografia por Emissão de Pósitrons
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
/
Prognostic_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
2008
Tipo de documento:
Article