[Shaping ability of two nickel-titanium rotary systems in simulated S-shaped canals].
Zhonghua Kou Qiang Yi Xue Za Zhi
; 43(1): 41-3, 2008 Jan.
Article
em Zh
| MEDLINE
| ID: mdl-18380974
OBJECTIVE: To evaluate the shaping ability of two nickel-titanium rotary systems (ProTaper and Hero642) in simulated S-shaped canals. METHODS: Thirty simulated S-shaped canals were randomly divided into three groups and prepared by ProTaper, Hero642, ProTaper combined with Hero642 respectively. All the canals were scanned before and after instrumentation, and the amount of material removed in the inner and outer wall and the canal width after instrumentation were measured with a computer image analysis program. RESULTS: There was significant difference in the amount of material removed at the inner side of apical curvature and outer side of apex between ProTaper combined with Hero642 and ProTaper files (P < 0.05) at the same tip size. The inner and outer wall of the canals were evenly prepared by ProTaper combined with Hero642, and the taper of canals were better than those prepared by Hero642. CONCLUSIONS: ProTaper combined with Hero 642 had better shaping ability to maintain the original shape and could create good taper canals in the simulated S-shaped canal model.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Titânio
/
Preparo de Canal Radicular
/
Cavidade Pulpar
/
Níquel
Tipo de estudo:
Evaluation_studies
Idioma:
Zh
Ano de publicação:
2008
Tipo de documento: