Your browser doesn't support javascript.
loading
Measurement of temporal resolution in dual source CT.
McCollough, Cynthia H; Schmidt, Bernhard; Yu, Lifeng; Primak, Andrew; Ulzheimer, Stefan; Bruder, Herbert; Flohr, Thomas G.
Afiliação
  • McCollough CH; Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905, USA. mccollough.cynthia@mayo.edu
Med Phys ; 35(2): 764-8, 2008 Feb.
Article em En | MEDLINE | ID: mdl-18383698
ABSTRACT
This work assesses the temporal resolution of dual-source computed tomography (CT) in a visually intuitive manner. Exploiting the principles of ring artifact creation, a phantom containing a highly attenuating delta function signal (a wire) was rotated at the same speed as the gantry, creating a partial ring artifact where the angular extent of the artifact provides a direct indication of the temporal resolution. A demonstration of the effect of the evaluated 165 and 83 ms nominal temporal resolutions on motion artifacts in cardiac CT is provided using patient data.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Artefatos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Artefatos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2008 Tipo de documento: Article