Improved attenuation correction via appropriate selection of respiratory-correlated PET data.
Daouk, Joël; Fin, Loïc; Bailly, Pascal; Meyer, Marc-Etienne.
Afiliação
  • Daouk J; Nuclear Medicine Department, Amiens University Medical Center, Amiens, France.
Comput Methods Programs Biomed ; 92(1): 90-8, 2008 Oct.
Article em En | MEDLINE | ID: mdl-18676054
ABSTRACT

PURPOSE:

We propose a respiratory-correlated PET data processing method (called "BH-CT-based") based on breath-hold CT acquisition to reduce the smearing effect and improve the attenuation correction. The resulting images are compared with the ungated PET images acquired using a standard, free-breathing clinical protocol.

METHODS:

The BH-CT-based method consisted of a list-mode acquisition with simultaneous respiratory signal recording. An additional breath-hold CT acquisition was also performed in order to define a tissue position from which PET events can be selected. A phantom study featured a 0.5-ml sphere (filled with 18F-fluorodeoxyglucose ((18)F-FDG) solution) pushed onto a rubber balloon (filled with (18)F-FDG solution and iodinated contrast agent). The feasibility of the BH-CT-based method was also assessed in two patients.

RESULTS:

In the phantom study, the contrast-to-noise ratios (CNRs) were -1.6 for the Ungated volume and 5.1 for the BH-CT-based volume. For patients, CNRs were higher for BH-CT-based volumes than those for Ungated volumes (17.3 vs. 6.3 and 7.3 vs. 3.8, for patients 1 and 2, respectively). Bias-variance measurements were performed and yielded bias reduction of 40% with BH-CT-based.

CONCLUSION:

The application of a BH-CT-based method decreases motion bias in PET images. This method resolves issues related to both PET-to-CT misregistration and erroneous attenuation correction and increases lesion detectability.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Tomografia Computadorizada por Raios X / Mecânica Respiratória / Tomografia por Emissão de Pósitrons / Técnicas de Imagem de Sincronização Respiratória / Neoplasias Tipo de estudo: Diagnostic_studies / Guideline Limite: Humans Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Tomografia Computadorizada por Raios X / Mecânica Respiratória / Tomografia por Emissão de Pósitrons / Técnicas de Imagem de Sincronização Respiratória / Neoplasias Tipo de estudo: Diagnostic_studies / Guideline Limite: Humans Idioma: En Ano de publicação: 2008 Tipo de documento: Article