Calibration of fundus images using spectral domain optical coherence tomography.
Lujan, Brandon J; Wang, Fenghua; Gregori, Giovanni; Rosenfeld, Philip J; Knighton, Robert W; Puliafito, Carmen A; Danis, Ronald P; Hubbard, Larry D; Chang, Robert T; Budenz, Donald L; Seider, Michael I; Knight, O'Rese.
Afiliação
  • Lujan BJ; Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida 33136, USA.
Ophthalmic Surg Lasers Imaging ; 39(4 Suppl): S15-20, 2008.
Article em En | MEDLINE | ID: mdl-18777875
ABSTRACT
BACKGROUND AND

OBJECTIVE:

Measurements performed on fundus images using current software are not accurate. Accurate measurements can be obtained only by calibrating a fundus camera using measurements between fixed retinal landmarks, such as the dimensions of the optic nerve, or by relying on a calibrated model eye provided by a reading center. However, calibrated spectral domain OCT (SD-OCT) could offer a convenient alternative method for the calibration of any fundus image. PATIENTS AND

METHODS:

The ability to measure exact distances on SD-OCT fundus images was tested by measuring the distance between the center of the fovea and the optic nerve. Calibrated SD-OCT scans measuring 6 X 6 X 2 mm centered on the fovea and the optic nerve were analyzed in 50 healthy right eyes. The foveal center was identified using cross-sectional SD-OCT images, and the center of the optic nerve was identified manually. The SD-OCT scans were registered to each other, and the distances between the center of the optic nerve and fovea were calculated. The overlay of these SD-OCT fundus images on photographic fundus images was performed.

RESULTS:

Any image of the fundus could be calibrated by overlaying the SD-OCT fundus image, and the measurements were consistent with previously defined calibration methods. The mean distance between the center of the fovea and the center of the optic nerve was 4.32 +/-0.32 mm. The line from the center of the optic nerve to the foveal center had a mean declination of 7.67 +/- 3.88 degrees. Mean horizontal displacement and vertical displacement were 4.27 +/- 0.29 mm and 0.58 +/- 0.29 mm, respectively.

CONCLUSIONS:

The overlay of the SD-OCT fundus image provides a convenient method for calibrating any image of the fundus. This approach should provide a uniform standard when comparing images from different devices and from different reading centers.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia de Coerência Óptica / Macula Lutea / Degeneração Macular Tipo de estudo: Prognostic_studies Limite: Aged80 / Female / Humans / Male Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia de Coerência Óptica / Macula Lutea / Degeneração Macular Tipo de estudo: Prognostic_studies Limite: Aged80 / Female / Humans / Male Idioma: En Ano de publicação: 2008 Tipo de documento: Article