Cost-effective diffuse reflectance spectroscopy device for quantifying tissue absorption and scattering in vivo.
J Biomed Opt
; 13(6): 060505, 2008.
Article
em En
| MEDLINE
| ID: mdl-19123646
A hybrid optical device that uses a multimode fiber coupled to a tunable light source for illumination and a 2.4-mm photodiode for detection in contact with the tissue surface is developed as a first step toward our goal of developing a cost-effective, miniature spectral imaging device to map tissue optical properties in vivo. This device coupled with an inverse Monte Carlo model of reflectance is demonstrated to accurately quantify tissue absorption and scattering in tissue-like turbid synthetic phantoms with a wide range of optical properties. The overall errors for quantifying the absorption and scattering coefficients are 6.0+/-5.6 and 6.1+/-4.7%, respectively. Compared with fiber-based detection, having the detector right at the tissue surface can significantly improve light collection efficiency, thus reducing the requirement for sophisticated detectors with high sensitivity, and this design can be easily expanded into a quantitative spectral imaging system for mapping tissue optical properties in vivo.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Fotometria
/
Radiometria
/
Análise Espectral
/
Iluminação
/
Interpretação de Imagem Assistida por Computador
/
Nefelometria e Turbidimetria
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
/
Health_economic_evaluation
/
Prognostic_studies
País como assunto:
America do norte
Idioma:
En
Ano de publicação:
2008
Tipo de documento:
Article