Micropumps operated by swelling and shrinking of temperature-sensitive hydrogels.
Lab Chip
; 9(4): 613-8, 2009 Feb 21.
Article
em En
| MEDLINE
| ID: mdl-19190798
This paper describes two types of polymeric micropumps based on the temperature-sensitive hydrogel poly(N-isopropylacrylamide). The gel actuators are realised as photopolymerised patterns and microgels. They are electrothermically controlled by resistive heating elements. The diffusion-based micropump contains a photopatterned monolithic actuator, which is placed within the pump chamber, and provides a valveless single layer set-up. The diffusion micropump is intended for low performance applications and can operate in two modes: peristaltic or pulsatile. The maximum operating parameters are a flow rate of 2.8 +/- 0.35 microl min(-1) and a back pressure of 1.28 kPa. The second type, a displacement pump, provides a higher performance (maximal 4.5 microl min(-1) and 15 kPa). The pump comprises a microgel-based actuator, which is placed within a separate actuator layer, and active microvalves. The specific features of the design and performance of the pumps are discussed.
Texto completo:
1
Base de dados:
MEDLINE
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Ano de publicação:
2009
Tipo de documento:
Article