Validation and compliance of a home monitoring device in pregnancy: microlife WatchBP home.
Chung, Yealin; de Greeff, Annemarie; Shennan, Andrew.
Afiliação
  • Chung Y; Maternal & Fetal Research Unit, King's College London, London, UK.
Hypertens Pregnancy ; 28(3): 348-59, 2009.
Article em En | MEDLINE | ID: mdl-19263287
ABSTRACT

OBJECTIVE:

To assess the accuracy and patient compliance in using a novel home blood pressure monitoring device in high-risk pregnancy.

METHODS:

Device accuracy was assessed according to the British Hypertension Society protocol in 45 pregnant women, including 15 with preeclampsia. Twenty-one high-risk pregnant women used the device in addition to their antenatal care.

RESULTS:

The device achieved a mean difference +/- SD of 0.4 +/- 7.3/-0.4 +/- 5.5 mmHg (pregnancy) and -2.6 +/- 7.0/0.8 +/- 4.4 mmHg (preeclampsia) for systolic/diastolic pressure. Eighty-one percent of women did at least 6 measurements/day and all women did at least 2 measurements/week.

CONCLUSION:

The Microlife WatchBP Home is accurate for use in pregnancy and increases surveillance in compliant patients.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pressão Sanguínea / Monitorização Ambulatorial da Pressão Arterial Limite: Adult / Female / Humans / Pregnancy Idioma: En Ano de publicação: 2009 Tipo de documento: Article