Your browser doesn't support javascript.
loading
Detection of lipid core coronary plaques in autopsy specimens with a novel catheter-based near-infrared spectroscopy system.
Gardner, Craig M; Tan, Huwei; Hull, Edward L; Lisauskas, Jennifer B; Sum, Stephen T; Meese, Thomas M; Jiang, Chunsheng; Madden, Sean P; Caplan, Jay D; Burke, Allen P; Virmani, Renu; Goldstein, James; Muller, James E.
Afiliação
  • Gardner CM; InfraReDx, Inc., Burlington, Massachusetts 01803, USA. cgardner@infraredx.com
JACC Cardiovasc Imaging ; 1(5): 638-48, 2008 Sep.
Article em En | MEDLINE | ID: mdl-19356494
ABSTRACT

OBJECTIVES:

This study sought to assess agreement between an intravascular near-infrared spectroscopy (NIRS) system and histology in coronary autopsy specimens.

BACKGROUND:

Lipid core plaques cannot be detected by conventional tests, yet are suspected to be the cause of most acute coronary syndromes. Near-infrared spectroscopy is widely used to determine the chemical content of substances. A NIRS system has been developed and used successfully in 99 patients.

METHODS:

Scanning NIRS was performed through blood in 212 coronary segments from 84 autopsy hearts. One histologic section was analyzed for every 2 mm of artery. Lipid core plaque of interest (LCP) was defined as a lipid core >60 degrees in circumferential extent, >200-microm thick, with a mean fibrous cap thickness <450 microm. The first 33 hearts were used to develop the algorithm; the subsequent 51 validation hearts were used in a prospective, double-blind manner to evaluate the accuracy of NIRS in detecting LCP. A NIRS-derived lipid core burden index for an entire artery was also validated by comparison to histologic findings.

RESULTS:

The LCPs were present in 115 of 2,649 (4.3%) sections from the 51 validation hearts. The algorithm prospectively identified LCP with a receiver-operator characteristic area of 0.80 (95% confidence interval [CI] 0.76 to 0.85). The lipid core burden index detected the presence or absence of any fibroatheroma with an area under the curve of 0.86 (95% CI 0.81 to 0.91). A retrospective analysis of lipid core burden index conducted in extreme artery segments with either no or extensive fibroatheroma yielded an area under the curve of 0.96 (95% CI 0.92 to 1.00), confirming the accuracy of spectroscopy in identifying plaques with markedly different lipid content under ideal circumstances.

CONCLUSIONS:

This novel catheter-based NIRS system accurately identified lipid core plaques through blood in a prospective study in coronary autopsy specimens. It is expected that this novel capability will be of assistance in the management of patients with coronary artery disease.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cateterismo / Espectroscopia de Luz Próxima ao Infravermelho / Vasos Coronários / Estenose Coronária / Lipídeos Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2008 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cateterismo / Espectroscopia de Luz Próxima ao Infravermelho / Vasos Coronários / Estenose Coronária / Lipídeos Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2008 Tipo de documento: Article