Testing the equivalence of several algorithms for calculation of coincidence summing corrections.
Appl Radiat Isot
; 67(10): 1952-6, 2009 Oct.
Article
em En
| MEDLINE
| ID: mdl-19647443
Five different numerical algorithms used for calculating coincidence summing corrections (CSCs) in HPGe gamma-spectrometry in case of complex radionuclide decay schemes were tested for their equivalence. Simulated point-source detection geometries and efficiency curves have been used to calculate and compare counting rates as well as CSCs of (133)Ba, (134)Cs and (152)Eu of 1MBq sources. Contributions of one X-ray per radionuclide were included. The algorithms were found to give identical results within their respective presuppositions.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2009
Tipo de documento:
Article