Your browser doesn't support javascript.
loading
Testing the equivalence of several algorithms for calculation of coincidence summing corrections.
Kanisch, G; Vidmar, T; Sima, O.
Afiliação
  • Kanisch G; Johann Heinrich v. Thünen-Institute, Hamburg, Germany. guenter.kanisch@vti.bund.de
Appl Radiat Isot ; 67(10): 1952-6, 2009 Oct.
Article em En | MEDLINE | ID: mdl-19647443
Five different numerical algorithms used for calculating coincidence summing corrections (CSCs) in HPGe gamma-spectrometry in case of complex radionuclide decay schemes were tested for their equivalence. Simulated point-source detection geometries and efficiency curves have been used to calculate and compare counting rates as well as CSCs of (133)Ba, (134)Cs and (152)Eu of 1MBq sources. Contributions of one X-ray per radionuclide were included. The algorithms were found to give identical results within their respective presuppositions.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2009 Tipo de documento: Article