Measurement delay associated with the Guardian RT continuous glucose monitoring system.
Diabet Med
; 27(1): 117-22, 2010 Jan.
Article
em En
| MEDLINE
| ID: mdl-20121899
ABSTRACT
AIMS:
Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D).METHODS:
Twelve children and adolescents with T1D treated by continuous subcutaneous insulin infusion (male/female 7/5; age 13.1 +/- 4.2 years; body mass index 21.9 +/- 4.3 kg/m(2); mean +/- sd) were studied over 19 h in a Clinical Research Facility. Guardian RT was calibrated every 6 h and sensor glucose measured every 5 min. Reference blood glucose was measured every 15 min using a YSI 2300 STAT Plus Analyser. A population compartment model of sensor glucose-blood glucose kinetics was adopted to estimate the time delay, the calibration scale and the calibration shift.RESULTS:
The population median of the time delay was 15.8 (interquartile range 15.2, 16.5) min, which was corroborated by correlation analysis between blood glucose and 15-min delayed sensor glucose. The delay has a relatively low intersubject variability, with 95% of individuals predicted to have delays between 10.4 and 24.3 min. Population medians (interquartile range) for the scale and shift are 0.800 (0.777, 0.823) (unitless) and 1.66 (1.47, 1.84) mmol/l, respectively.CONCLUSIONS:
In young subjects with T1D, the total time delay associated with the Guardian RT system was approximately 15 min. This is twice that expected on physiological grounds, suggesting a 5- to 10-min delay because of data processing. Delays above 25 min are rarely to be observed.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Glicemia
/
Automonitorização da Glicemia
/
Diabetes Mellitus Tipo 1
Tipo de estudo:
Prognostic_studies
/
Risk_factors_studies
Limite:
Adolescent
/
Child
/
Female
/
Humans
/
Male
Idioma:
En
Ano de publicação:
2010
Tipo de documento:
Article