Measurement delay associated with the Guardian RT continuous glucose monitoring system.
Wei, C; Lunn, D J; Acerini, C L; Allen, J M; Larsen, A M; Wilinska, M E; Dunger, D B; Hovorka, R.
Afiliação
  • Wei C; MRC Biostatistics Unit, Institute of Public Health, University of Cambridge, Cambridge, UK. chen.wei@mrc-bsu.cam.ac.uk
Diabet Med ; 27(1): 117-22, 2010 Jan.
Article em En | MEDLINE | ID: mdl-20121899
ABSTRACT

AIMS:

Using compartment modelling, we assessed the time delay between blood glucose and sensor glucose measured by the Guardian RT continuous glucose monitoring system in young subjects with Type 1 diabetes (T1D).

METHODS:

Twelve children and adolescents with T1D treated by continuous subcutaneous insulin infusion (male/female 7/5; age 13.1 +/- 4.2 years; body mass index 21.9 +/- 4.3 kg/m(2); mean +/- sd) were studied over 19 h in a Clinical Research Facility. Guardian RT was calibrated every 6 h and sensor glucose measured every 5 min. Reference blood glucose was measured every 15 min using a YSI 2300 STAT Plus Analyser. A population compartment model of sensor glucose-blood glucose kinetics was adopted to estimate the time delay, the calibration scale and the calibration shift.

RESULTS:

The population median of the time delay was 15.8 (interquartile range 15.2, 16.5) min, which was corroborated by correlation analysis between blood glucose and 15-min delayed sensor glucose. The delay has a relatively low intersubject variability, with 95% of individuals predicted to have delays between 10.4 and 24.3 min. Population medians (interquartile range) for the scale and shift are 0.800 (0.777, 0.823) (unitless) and 1.66 (1.47, 1.84) mmol/l, respectively.

CONCLUSIONS:

In young subjects with T1D, the total time delay associated with the Guardian RT system was approximately 15 min. This is twice that expected on physiological grounds, suggesting a 5- to 10-min delay because of data processing. Delays above 25 min are rarely to be observed.