Your browser doesn't support javascript.
loading
Defibrillation threshold decrease with the supradiaphragmatic extracardiac implantable cardioverter-defibrillator implantation technique.
Berruezo, Antonio; Zeljko, Hrvojka M; Bartrons, Joaquim; Mayol, Javier; Prada, Fredy; Brugada, Josep.
Afiliação
  • Berruezo A; Arrhythmia Section, Cardiology Department, Thorax Institute, Hospital Clinic, Barcelona, Catalonia, Spain. berruezo@clinic.ub.es
Europace ; 12(11): 1649-51, 2010 Nov.
Article em En | MEDLINE | ID: mdl-20543197
Despite advances in implantable cardioverter-defibrillator (ICD) technology, the optimal ICD implantation technique for pediatric patients has not yet been established. One increasingly used option is totally extracardiac implantation. However, concern exists about the high defibrillation threshold (DFT) at the moment of implantation or during follow-up. We report the case of a 3-year-old boy with repetitive syncopal idiopathic ventricular tachycardia episodes treated with ICD implantation using the extracardiac technique. Changing device position from abdominal to a supradiaphragmatic, solved unsafe elevated discharge impedance and DFT during follow-up.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Síncope / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Diagnostic_studies Limite: Child, preschool / Humans / Male Idioma: En Ano de publicação: 2010 Tipo de documento: Article