Your browser doesn't support javascript.
loading
[Evaluation of capability of cell count and detection of tumor cells in cerebrospinal and body fluids by automated hematology analyzer].
Takemura, Hiroyuki; Tabe, Yoko; Ishii, Kiyoshi; Kobayashi, Yuki; Kuno, Yutaka; Horii, Takashi; Miyake, Kazunori; Miida, Takashi; Ohsaka, Akimichi.
Afiliação
  • Takemura H; Department of Clinical Laboratory Medicine, Juntendo University School of Medicine, Bunkyo-ku, Tokyo 113-8421, Japan.
Rinsho Byori ; 58(6): 559-64, 2010 Jun.
Article em Ja | MEDLINE | ID: mdl-20662266
ABSTRACT
Sysmex XE-5000 offers the body fluid modus which provides the opportunity to count and differentiate leukocytes in body fluids and cerebrospinal fluid (CFS). In this study, we evaluated the basic performance of this application using routinely obtained samples in comparison with manual counting. Reproducibility study yielded good results in samples with a high white blood cell (WBC) count, whereas relatively high imprecision was observed at low WBC counts. Linearity was established up to 1,500 cells/microL in CFS and 5,600 cells/microL in body fluid. The cell count by XE-5000 was highly correlated with that of the microscopic reference method. Highly fluorescent body fluid cells percent (HF-BF%) was observed in samples with tumor cells or activated macrophages, which provides information about the possible presence of tumor cells. In conclusion, total and differential WBC counts in body fluid and CFS can be reliably determined by XE-5000 in samples with increased cell counts. XE-5000 also provides screening information about the presence of tumor cells for further manual examination.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Líquidos Corporais / Contagem de Células / Líquido Cefalorraquidiano / Citodiagnóstico / Hematologia / Células Neoplásicas Circulantes Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: Ja Ano de publicação: 2010 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Líquidos Corporais / Contagem de Células / Líquido Cefalorraquidiano / Citodiagnóstico / Hematologia / Células Neoplásicas Circulantes Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: Ja Ano de publicação: 2010 Tipo de documento: Article