International reproducibility of single breathhold T2* MR for cardiac and liver iron assessment among five thalassemia centers.
J Magn Reson Imaging
; 32(2): 315-9, 2010 Aug.
Article
em En
| MEDLINE
| ID: mdl-20677256
PURPOSE: To examine the reproducibility of the single breathhold T2* technique from different scanners, after installation of standard methodology in five international centers. MATERIALS AND METHODS: Up to 10 patients from each center were scanned twice locally for local interstudy reproducibility of heart and liver T2*, and then flown to a central MR facility to be rescanned on a reference scanner for intercenter reproducibility. Interobserver reproducibility for all scans was also assessed. RESULTS: Of the 49 patients scanned, the intercenter reproducibility for T2* was 5.9% for the heart and 5.8% for the liver. Local interstudy reproducibility for T2* was 7.4% for the heart and 4.6% for the liver. Interobserver reproducibility for T2* was 5.4% for the heart and 4.4% for the liver. CONCLUSION: These data indicate that T2* MR may be developed into a widespread test for tissue siderosis providing that well-defined and approved imaging and analysis techniques are used.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Talassemia
/
Imageamento por Ressonância Magnética
/
Sobrecarga de Ferro
/
Ferro
/
Fígado
/
Miocárdio
Tipo de estudo:
Clinical_trials
Limite:
Adult
/
Female
/
Humans
/
Male
Idioma:
En
Ano de publicação:
2010
Tipo de documento:
Article