Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry.
Poole, Jeanne E; Gleva, Marye J; Mela, Theofanie; Chung, Mina K; Uslan, Daniel Z; Borge, Richard; Gottipaty, Venkateshwar; Shinn, Timothy; Dan, Dan; Feldman, Leon A; Seide, Hanscy; Winston, Stuart A; Gallagher, John J; Langberg, Jonathan J; Mitchell, Kevin; Holcomb, Richard.
Afiliação
  • Poole JE; University of Washington, Division of Cardiology, 1959 NE Pacific Street, Seattle, WA 98198-6422, USA. jpoole@u.washington.edu
Circulation ; 122(16): 1553-61, 2010 Oct 19.
Article em En | MEDLINE | ID: mdl-20921437
ABSTRACT

BACKGROUND:

Prospective studies defining the risk associated with pacemaker or implantable cardioverter-defibrillator replacement surgeries do not exist. These procedures are generally considered low risk despite results from recent retrospective series reporting higher rates. METHODS AND

RESULTS:

We prospectively assessed predefined procedure-related complication rates associated with elective pacemaker or implantable cardioverter-defibrillator generator replacements over 6 months of follow-up. Two groups were studied those without (cohort 1) and those with (cohort 2) a planned transvenous lead addition for replacement or upgrade to a device capable of additional therapies. Complications were adjudicated by an independent events committee. Seventy-two US academic and private practice centers participated. Major complications occurred in 4.0% (95% confidence interval, 2.9 to 5.4) of 1031 cohort 1 patients and 15.3% (95% confidence interval, 12.7 to 18.1) of 713 cohort 2 patients. In both cohorts, major complications were higher with implantable cardioverter-defibrillator compared with pacemaker generator replacements. Complications were highest in patients who had an upgrade to or a revised cardiac resynchronization therapy device (18.7%; 95% confidence interval, 15.1 to 22.6). No periprocedural deaths occurred in either cohort, although 8 later procedure-related deaths occurred in cohort 2. The 6-month infection rates were 1.4% (95% confidence interval, 0.7 to 2.3) and 1.1% (95% confidence interval, 0.5 to 2.2) for cohorts 1 and 2, respectively.

CONCLUSIONS:

Pacemaker and implantable cardioverter-defibrillator generator replacements are associated with a notable complication risk, particularly those with lead additions. These data support careful decision making before device replacement, when managing device advisories, and when considering upgrades to more complex systems.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Complicações Pós-Operatórias / Sistema de Registros / Desfibriladores Implantáveis / Remoção de Dispositivo Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Complicações Pós-Operatórias / Sistema de Registros / Desfibriladores Implantáveis / Remoção de Dispositivo Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2010 Tipo de documento: Article