Commercial software upgrades may significantly alter Perfusion CT parameter values in colorectal cancer.
Goh, Vicky; Shastry, Manu; Engledow, Alec; Reston, Jonathan; Wellsted, David M; Peck, Jacqui; Endozo, Raymondo; Rodriguez-Justo, Manuel; Taylor, Stuart A; Halligan, Steve; Groves, Ashley M.
Afiliação
  • Goh V; Paul Strickland Scanner Centre, Mount Vernon Hospital, Rickmansworth Road, Northwood, Middlesex HA6 2RN, UK. vicky.goh@stricklandscanner.org.uk
Eur Radiol ; 21(4): 744-9, 2011 Apr.
Article em En | MEDLINE | ID: mdl-20922392
ABSTRACT

OBJECTIVE:

To determine how commercial software platform upgrades impact on derived parameters for colorectal cancer. MATERIALS AND

METHODS:

Following ethical approval, 30 patients with suspected colorectal cancer underwent Perfusion CT using integrated 64 detector PET/CT before surgery. Analysis was performed using software based on modified distributed parameter analysis (Perfusion software version 4; Perfusion 4.0), then repeated using the previous version (Perfusion software version 3; Perfusion 3.0). Tumour blood flow (BF), blood volume (BV), mean transit time (MTT) and permeability surface area product (PS) were determined for identical regions-of-interest. Slice-by-slice and 'whole tumour' variance was assessed by Bland-Altman analysis.

RESULTS:

Mean BF, BV and PS was 20.4%, 59.5%, and 106% higher, and MTT 14.3% shorter for Perfusion 4.0 than Perfusion 3.0. The mean difference (95% limits of agreement) were +13.5 (-44.9 to 72.0), +2.61 (-0.06 to 5.28), -1.23 (-6.83 to 4.36), and +14.2 (-4.43 to 32.8) for BF, BV, MTT and PS respectively. Within subject coefficient of variation was 36.6%, 38.0%, 27.4% and 60.6% for BF, BV, MTT and PS respectively indicating moderate to poor agreement.

CONCLUSION:

Software version upgrades of the same software platform may result in significantly different parameter values, requiring adjustments for cross-version comparison.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Retais / Interpretação de Imagem Radiográfica Assistida por Computador / Neoplasias Colorretais / Adenocarcinoma / Tomografia Computadorizada por Raios X / Neoplasias do Colo Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Retais / Interpretação de Imagem Radiográfica Assistida por Computador / Neoplasias Colorretais / Adenocarcinoma / Tomografia Computadorizada por Raios X / Neoplasias do Colo Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article