Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients.
De Backer, Daniel; Marx, Gernot; Tan, Andrew; Junker, Christopher; Van Nuffelen, Marc; Hüter, Lars; Ching, Willy; Michard, Frédéric; Vincent, Jean-Louis.
Afiliação
  • De Backer D; Department of Intensive Care, Erasme University Hospital, Université Libre de Bruxelles, Route de Lennik 808, 1070 Brussels, Belgium. ddebacke@ulb.ac.be
Intensive Care Med ; 37(2): 233-40, 2011 Feb.
Article em En | MEDLINE | ID: mdl-21153399
ABSTRACT

PURPOSE:

Second-generation FloTrac software has been shown to reliably measure cardiac output (CO) in cardiac surgical patients. However, concerns have been raised regarding its accuracy in vasoplegic states. The aim of the present multicenter study was to investigate the accuracy of the third-generation software in patients with sepsis, particularly when total systemic vascular resistance (TSVR) is low.

METHODS:

Fifty-eight septic patients were included in this prospective observational study in four university-affiliated ICUs. Reference CO was measured by bolus pulmonary thermodilution (iCO) using 3-5 cold saline boluses. Simultaneously, CO was computed from the arterial pressure curve recorded on a computer using the second-generation (CO(G2)) and third-generation (CO(G3)) FloTrac software. CO was also measured by semi-continuous pulmonary thermodilution (CCO).

RESULTS:

A total of 401 simultaneous measurements of iCO, CO(G2), CO(G3), and CCO were recorded. The mean (95%CI) biases between CO(G2) and iCO, CO(G3) and iCO, and CCO and iCO were -10 (-15 to -5)% [-0.8 (-1.1 to -0.4) L/min], 0 (-4 to 4)% [0 (-0.3 to 0.3) L/min], and 9 (6-13)% [0.7 (0.5-1.0) L/min], respectively. The percentage errors were 29 (20-37)% for CO(G2), 30 (24-37)% for CO(G3), and 28 (22-34)% for CCO. The difference between iCO and CO(G2) was significantly correlated with TSVR (r(2) = 0.37, p < 0.0001). A very weak (r(2) = 0.05) relationship was also observed for the difference between iCO and CO(G3).

CONCLUSIONS:

In patients with sepsis, the third-generation FloTrac software is more accurate, as precise, and less influenced by TSVR than the second-generation software.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pressão Sanguínea / Software / Cateterismo de Swan-Ganz / Débito Cardíaco / Sepse / Monitorização Fisiológica Tipo de estudo: Clinical_trials / Observational_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pressão Sanguínea / Software / Cateterismo de Swan-Ganz / Débito Cardíaco / Sepse / Monitorização Fisiológica Tipo de estudo: Clinical_trials / Observational_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2011 Tipo de documento: Article