Radiation damage in room-temperature data acquisition with the PILATUS 6M pixel detector.
Rajendran, Chitra; Dworkowski, Florian S N; Wang, Meitian; Schulze-Briese, Clemens.
Afiliação
  • Rajendran C; Swiss Light Source at Paul Scherrer Institute, CH-5232 Villigen, Switzerland. chitra.rajendran@psi.ch
J Synchrotron Radiat ; 18(Pt 3): 318-28, 2011 May.
Article em En | MEDLINE | ID: mdl-21525639
The first study of room-temperature macromolecular crystallography data acquisition with a silicon pixel detector is presented, where the data are collected in continuous sample rotation mode, with millisecond read-out time and no read-out noise. Several successive datasets were collected sequentially from single test crystals of thaumatin and insulin. The dose rate ranged between ∼ 1320 Gy s(-1) and ∼ 8420 Gy s(-1) with corresponding frame rates between 1.565 Hz and 12.5 Hz. The data were analysed for global radiation damage. A previously unreported negative dose-rate effect is observed in the indicators of global radiation damage, which showed an approximately 75% decrease in D(1/2) at sixfold higher dose rate. The integrated intensity decreases in an exponential manner. Sample heating that could give rise to the enhanced radiation sensitivity at higher dose rate is investigated by collecting data between crystal temperatures of 298 K and 353 K. UV-Vis spectroscopy is used to demonstrate that disulfide radicals and trapped electrons do not accumulate at high dose rates in continuous data collection.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2011 Tipo de documento: Article