The Integrity bare-metal stent made by continuous sinusoid technology.
Turco, Mark A.
Afiliação
  • Turco MA; Center for Cardiac and Vascular Research, Washington Adventist Hospital, 7600 Carroll Ave., Takoma Park, MD 20912-6367, USA. mturco@ahm.com
Expert Rev Med Devices ; 8(3): 303-6, 2011 May.
Article em En | MEDLINE | ID: mdl-21542702
The Integrity Coronary Stent System (Medtronic Vascular, CA, USA) is a low-profile, open-cell, cobalt-chromium-alloy advanced bare-metal iteration of the well-known Driver/Micro-Driver Coronary Stent System (Medtronic Vascular). The Integrity stent is made with a process called continuous sinusoid technology. This process allows stent construction via wrapping a single thin strand of wire around a mandrel in a sinusoid configuration, with laser fusion of adjacent crowns. The wire-forming process and fusion pattern provide the stent with a continuous preferential bending plane, intended to allow easier access to, and smoother tracking within, distal and tortuous vessels while radial strength is maintained. Continuous sinusoid technology represents innovation in the design of stent platforms and will provide a future stent platform for newer technology, including drug-eluting stent platforms, drug-filled stents and core wire stents.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Stents Limite: Humans Idioma: En Ano de publicação: 2011 Tipo de documento: Article