The Integrity bare-metal stent made by continuous sinusoid technology.
Expert Rev Med Devices
; 8(3): 303-6, 2011 May.
Article
em En
| MEDLINE
| ID: mdl-21542702
The Integrity Coronary Stent System (Medtronic Vascular, CA, USA) is a low-profile, open-cell, cobalt-chromium-alloy advanced bare-metal iteration of the well-known Driver/Micro-Driver Coronary Stent System (Medtronic Vascular). The Integrity stent is made with a process called continuous sinusoid technology. This process allows stent construction via wrapping a single thin strand of wire around a mandrel in a sinusoid configuration, with laser fusion of adjacent crowns. The wire-forming process and fusion pattern provide the stent with a continuous preferential bending plane, intended to allow easier access to, and smoother tracking within, distal and tortuous vessels while radial strength is maintained. Continuous sinusoid technology represents innovation in the design of stent platforms and will provide a future stent platform for newer technology, including drug-eluting stent platforms, drug-filled stents and core wire stents.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Desenho de Prótese
/
Stents
Limite:
Humans
Idioma:
En
Ano de publicação:
2011
Tipo de documento:
Article