Retrospective analyses of complications associated with transcutaneous replacement of percutaneous gastrostomy and jejunostomy feeding devices.
Nishiwaki, Shinji; Araki, Hiroshi; Fang, John C; Hayashi, Motoshi; Takada, Jun; Iwashita, Masahide; Tagami, Atsushi; Hatakeyama, Hiroo; Hayashi, Takao; Maeda, Teruo; Saito, Koshiro.
Afiliação
  • Nishiwaki S; Department of Internal Medicine, Nishimino Kosei Hospital, Yoro-gun, Gifu, Japan.
Gastrointest Endosc ; 74(4): 784-91, 2011 Oct.
Article em En | MEDLINE | ID: mdl-21802678
ABSTRACT

BACKGROUND:

Feeding device replacement is often required for long-term maintenance after initial percutaneous endoscopic gastrostomy or jejunostomy placement. Although there are several case reports on serious complications of gastrostomy device replacement, there are few reports of an overall analysis of the complications associated with feeding device replacement.

OBJECTIVE:

To evaluate the frequency and variety of complications of transcutaneous replacement of feeding devices.

DESIGN:

A retrospective study.

SETTING:

Single center Nishimino Kosei Hospital. PATIENTS This study involved 363 consecutive patients undergoing a total of 1265 percutaneous gastrostomy or jejunostomy device replacements from March 2000 to September 2010. INTERVENTION A new replacement device was inserted through the ostomy tract by using an obturator after traction removal of the previous device. Endoscopic treatments were performed in the cases of fistula disruption or hemorrhage. MAIN OUTCOME MEASUREMENTS Complications and their outcomes.

RESULTS:

Gastrostomy and jejunostomy devices were replaced 1126 and 139 times, respectively. There were 16 complications (1.3% of total replacements) consisting of 10 cases of fistula disruption caused by misplacement of replacement devices into the peritoneal cavity, 4 cases of hemorrhage, and 1 case each of colocutaneous fistula and device breakage. Anticoagulation or antiplatelet medications were continued in all 4 hemorrhage cases but in only 27 of 347 (7.7%) complication-free cases (P < .0001). There were no replacement-related adverse events that required surgical repair.

LIMITATIONS:

A single center, retrospective analysis.

CONCLUSION:

Fistula disruption and hemorrhage were the most common complications associated with device replacement. In patients on anticoagulants, caution is necessary to avoid hemorrhage after replacement. It is also important to verify that the replaced device is located in the GI tract lumen before feeding.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Jejunostomia / Gastrostomia / Endoscopia Gastrointestinal / Nutrição Enteral / Remoção de Dispositivo Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Jejunostomia / Gastrostomia / Endoscopia Gastrointestinal / Nutrição Enteral / Remoção de Dispositivo Tipo de estudo: Etiology_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article