Opacification of intraocular lenses implanted during infancy: a clinicopathologic study of 4 explanted intraocular lenses.
Pehere, Niranjan K; Bojja, Sreedhar; Vemuganti, Geeta K; Vaddavalli, Pravin K; Samant, Monica; Jalali, Subhadra; Bhate, Manjushree.
Afiliação
  • Pehere NK; The David Brown Children's Eye Care Center, L.V. Prasad Eye Institute, Vijayawada, India.
Ophthalmology ; 118(11): 2128-2132.e1, 2011 Nov.
Article em En | MEDLINE | ID: mdl-21862135
ABSTRACT

OBJECTIVE:

To report the clinicopathologic features of 4 opacified, single-piece, hydrophilic, acrylic intraocular lenses (IOLs) explanted from children who had undergone IOL implantation during infancy.

DESIGN:

Observational case series.

PARTICIPANTS:

Four IOLs explanted from 4 eyes of 3 children for visually significant opacification were included in the study.

METHODS:

The clinical details of each case were obtained to look for possible risk factors for IOL opacification. The explanted IOLs were subjected to gross examination, staining by alizarin red 1% for calcium, scanning electron microscopy, and energy-dispersive x-ray spectroscopy (EDS). Levels of calcium and phosphorous were analyzed in the serum of all cases and in the aqueous humor of 1 case. MAIN OUTCOME

MEASURES:

Morphologic features and composition of deposits.

RESULTS:

Two cases had congenital cataract while one case was after bilateral lens sparing vitrectomy for retinopathy of prematurity. All underwent surgery during infancy with implantation of an IOL. The IOLs were explanted 8 months after surgery from 4 eyes of 3 children at the age of 17, 25, and 26. All the children received a single-piece hydrophilic acrylic IOL. The IOLs were in situ for an average duration of 13.86 months. The deposits were in the shape of a bicycle wheel on 3 IOLs and looked like fish eggs on 1 IOL. All deposits stained bright orange with alizarin red. On EDS, the deposits were found to be composed of calcium, phosphate, and silicone.

CONCLUSIONS:

The morphologic features and composition of IOL deposits in 2 cases were similar to those of earlier reports in adults. The hydrophilic nature of the IOL material, sulcus implantation, and postoperative inflammation may be possible risk factors for opacification. FINANCIAL DISCLOSURE(S) Proprietary or commercial disclosure may be found after the references.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Falha de Prótese / Extração de Catarata / Implante de Lente Intraocular / Lentes Intraoculares Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans / Infant / Newborn Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Falha de Prótese / Extração de Catarata / Implante de Lente Intraocular / Lentes Intraoculares Tipo de estudo: Etiology_studies / Risk_factors_studies Limite: Humans / Infant / Newborn Idioma: En Ano de publicação: 2011 Tipo de documento: Article