Percutaneous implantation of the Edwards SAPIEN transcatheter heart valve for conduit failure in the pulmonary position: early phase 1 results from an international multicenter clinical trial.
Kenny, Damien; Hijazi, Ziyad M; Kar, Saibal; Rhodes, John; Mullen, Michael; Makkar, Raj; Shirali, Girish; Fogel, Mark; Fahey, John; Heitschmidt, Mary G; Cain, Christopher.
Afiliação
  • Kenny D; Rush University Medical Center, 1653 West Congress Parkway, Chicago, IL 60612, USA.
J Am Coll Cardiol ; 58(21): 2248-56, 2011 Nov 15.
Article em En | MEDLINE | ID: mdl-22078433
ABSTRACT

OBJECTIVES:

The purpose of this study was to evaluate the safety and effectiveness of the Edwards SAPIEN transcatheter heart valve (Edwards Lifesciences LLC, Irvine, California) in the pulmonary position in patients with moderate to severe pulmonary regurgitation with or without stenosis.

BACKGROUND:

Transcatheter pulmonary valve replacement is evolving, but to date, experience has been limited to the Melody valve (Medtronic Inc., Minneapolis, Minnesota).

METHODS:

Eligible patients with dysfunctional right ventricle-to-pulmonary artery conduits were screened if body weight was ≥35 kg and the in situ conduit diameter was ≥16 mm and ≤24 mm. Standardized implantation and follow-up protocols were used.

RESULTS:

Thirty-six patients from 4 centers were recruited between April 2008 and May 2010. Mean body weight was 73.4 ± 22.9 kg. Successful valve deployment was achieved in 33 of 34 attempts (97.1%). Valve migration occurred in 3 patients, with 2 requiring surgical retrieval; however, 1 patient underwent successful perventricular valve implantation. Further intraprocedure complications included pulmonary hemorrhage (n = 2), ventricular fibrillation (n = 1), and stent migration (n = 1). Pullback gradient across the conduit decreased from 26.8 ± 18.4 mm Hg to 11.7 ± 8.0 mm Hg (p < 0.001). The right ventricular/aortic pressure ratio decreased from 0.6 ± 0.2 to 0.4 ± 0.1 (p < 0.001). Peak Doppler gradient across the right ventricular outflow tract decreased from 41.9 ± 27.9 mm Hg to 19.1 ± 13.3 mm Hg (p < 0.001). At 6-month follow-up, all patients were alive. The number of patients with New York Heart Association functional class I increased from 5 at baseline to 27 at follow-up. Pulmonary regurgitation was ≤2+ in 97% of patients. Freedom from reintervention was 97% with 1 patient undergoing elective placement of a second valve due to conduit-induced distortion of the initial implant.

CONCLUSIONS:

Transcatheter pulmonary valve replacement using the Edwards SAPIEN transcatheter heart valve is safe and effective in patients with dysfunctional right ventricle-to-pulmonary artery conduits.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Valva Pulmonar / Insuficiência da Valva Pulmonar / Bioprótese / Obstrução do Fluxo Ventricular Externo / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Valva Pulmonar / Insuficiência da Valva Pulmonar / Bioprótese / Obstrução do Fluxo Ventricular Externo / Cateterismo Cardíaco / Implante de Prótese de Valva Cardíaca Tipo de estudo: Clinical_trials / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2011 Tipo de documento: Article