Your browser doesn't support javascript.
loading
Feasibility, reproducibility, and reliability for the T*2 iron evaluation at 3 T in comparison with 1.5 T.
Magn Reson Med ; 68(2): 543-51, 2012 Aug.
Article em En | MEDLINE | ID: mdl-22127999
ABSTRACT
This study aimed to determine the feasibility, reproducibility, and reliability of the multiecho T*(2) Magnetic resonance imaging technique at 3 T for myocardial and liver iron burden quantification and the relationship between T*(2) values at 3 and 1.5 T. Thirty-eight transfusion-dependent patients and 20 healthy subjects were studied. Cardiac segmental and global T*(2) values were calculated after developing a correction map to compensate the artifactual T*(2) variations. The hepatic T*(2) value was determined over a region of interest. The intraoperator and interoperator reproducibility for T*(2) measurements at 3 T was good. A linear relationship was found between patients' R *2 (1000/T*(2) ) values at 3 and 1.5 T. Segmental correction factors were significantly higher at 3 T. A conversion formula returning T*(2) values at 1.5 T from values at 3 T was proposed. A good diagnostic reliability for T*(2) assessment at 3 T was demonstrated. Lower limits of normal for 3 T T*(2) values were 23.3 ms, 21.1 ms, and 11.7 ms, for the global heart, mid-ventricular septum, and liver, respectively. In conclusion, T*(2) quantification of iron burden in the mid-ventricular septum, global heart, and no heavy-moderate livers resulted to be feasible, reproducible, and reliable at 3 T. Segmental heart T*(2) analysis at 3 T may be challenging due to significantly higher susceptibility artifacts.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Talassemia / Imageamento por Ressonância Magnética / Aumento da Imagem / Sobrecarga de Ferro Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Talassemia / Imageamento por Ressonância Magnética / Aumento da Imagem / Sobrecarga de Ferro Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2012 Tipo de documento: Article