A 24 keV liquid-metal-jet x-ray source for biomedical applications.
Larsson, D H; Takman, P A C; Lundström, U; Burvall, A; Hertz, H M.
Afiliação
  • Larsson DH; Biomedical and X-Ray Physics, Department of Applied Physics, Royal Institute of Technology/Albanova, SE-10691 Stockholm, Sweden. daniel.larsson@biox.kth.se
Rev Sci Instrum ; 82(12): 123701, 2011 Dec.
Article em En | MEDLINE | ID: mdl-22225218
We present a high-brightness 24-keV electron-impact microfocus x-ray source based on continuous operation of a heated liquid-indium/gallium-jet anode. The 30-70 W electron beam is magnetically focused onto the jet, producing a circular 7-13 µm full width half maximum x-ray spot. The measured spectral brightness at the 24.2 keV In K(α) line is 3 × 10(9) photons∕(s × mm(2) × mrad(2) × 0.1% BW) at 30 W electron-beam power. The high photon energy compared to existing liquid-metal-jet sources increases the penetration depth and allows imaging of thicker samples. The applicability of the source in the biomedical field is demonstrated by high-resolution imaging of a mammography phantom and a phase-contrast angiography phantom.