Noninvasive continuous arterial blood pressure monitoring with Nexfin®.
Martina, Jerson R; Westerhof, Berend E; van Goudoever, Jeroen; de Beaumont, Edouard M F H; Truijen, Jasper; Kim, Yu-Sok; Immink, Rogier V; Jöbsis, Dorothea A; Hollmann, Markus W; Lahpor, Jaap R; de Mol, Bas A J M; van Lieshout, Johannes J.
Afiliação
  • Martina JR; Department of Cardiothoracic Surgery, University Medical Center Utrecht, Utrecht, The Netherlands.
Anesthesiology ; 116(5): 1092-103, 2012 May.
Article em En | MEDLINE | ID: mdl-22415387
ABSTRACT

BACKGROUND:

If invasive measurement of arterial blood pressure is not warranted, finger cuff technology can provide continuous and noninvasive monitoring. Finger and radial artery pressures differ; Nexfin® (BMEYE, Amsterdam, The Netherlands) measures finger arterial pressure and uses physiologic reconstruction methodologies to obtain values comparable to invasive pressures.

METHODS:

Intra-arterial pressure (IAP) and noninvasive Nexfin arterial pressure (NAP) were measured in cardiothoracic surgery patients, because invasive pressures are available. NAP-IAP differences were analyzed during 30 min. Tracking was quantified by within-subject precision (SD of individual NAP-IAP differences) and correlation coefficients. The ranges of pressure change were quantified by within-subject variability (SD of individual averages of NAP and IAP). Accuracy and precision were expressed as group average ± SD of the differences and considered acceptable when smaller than 5 ± 8 mmHg, the Association for the Advancement of Medical Instrumentation criteria.

RESULTS:

NAP and IAP were obtained in 50 (34-83 yr, 40 men) patients. For systolic, diastolic, mean arterial, and pulse pressure, median (25-75 percentiles) correlation coefficients were 0.96 (0.91-0.98), 0.93 (0.87-0.96), 0.96 (0.90-0.97), and 0.94 (0.85-0.98), respectively. Within-subject precisions were 4 ± 2, 3 ± 1, 3 ± 2, and 3 ± 2 mmHg, and within-subject variations 13 ± 6, 6 ± 3, 9 ± 4, and 7 ± 4 mmHg, indicating precision over a wide range of pressures. Group average ± SD of the NAP-IAP differences were -1 ± 7, 3 ± 6, 2 ± 6, and -3 ± 4 mmHg, meeting criteria. Differences were not related to mean arterial pressure or heart rate.

CONCLUSION:

Arterial blood pressure can be measured noninvasively and continuously using physiologic pressure reconstruction. Changes in pressure can be followed and values are comparable to invasive monitoring.